61 77 476 D 3495
50
D 3495
.
201.01.02
.
02
08.2007
12
Electromagnetic compatibility
Sirona Dental Systems GmbH
Interference immunity
Operating Instructions and Installation
XIOS USB
The customer or user of the
UNIT
should make sure that it is used in such an
environment.
Interference
immunity tests
IEC 60601-1-2 test level Compliance level
Electromagnetic environment –
guidelines
Electrostatic
discharge (ESD)
according to
IEC
61000-4-2
± 6 kV contact discharge
± 8 kV air discharge
± 6 kV contact discharge
± 8 kV air discharge
Floors should be made of wood or
concrete or finished with ceramic tiling.
If floors are covered with synthetic
material, the relative humidity must be
at least 30%.
Electrical fast
transient/burst
according to
IEC 61000-4-4
± 1kV for input and output
lines
± 2 kV for power supply
lines
± 1 kV for input and
output lines
± 2 kV for power supply
lines
The quality of the line power supply
should be that of a typical commercial
or hospital environment.
Surge voltages
according to
IEC
61000-4-5
± 1 kV differential mode
± 2 kV common mode
voltage
± 1 kV differential mode
± 2 kV common mode
voltage
The quality of the line power supply
should be that of a typical commercial
or hospital environment.
Voltage dips, short
interruptions and
variations of the
power supply
according to
IEC
61000-4-11
<5% U
T
for ½ period (>95%
dip of U
T
)
40% UT for 5 periods (60%
dip of U
T
)
70% UT for 25 periods (30%
dip of U
T
)
<5% UT for 5 sec. (>95%
dip of U
T
<5% U
T
for ½ period
(>95% dip of U
T
)
40% U
T
for 5 periods
(60% dip of U
T
)
70 % U
T
for 25 periods
(30% dip of U
T
)
<5% U
T
for 5sec. (>95%
dip of U
T
The quality of the line power supply
should be that of a typical commercial
or hospital environment.
If the user of the
UNIT
requires it to
continue functioning following
interruptions of the power supply, it is
recommended to have the
UNIT
powered by an uninterruptible power
supply or a battery.
Magnetic field of
power frequencies
(50/60 Hz) according
to
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Remarks: U
T
is the AC supply voltage prior to application of the test level.