Electromagnetic compatibility
Operating Instructions and Installation
XIOS USB
Accessories
12
Electromagnetic compatibility
12.1
Accessories
z
The UNIT may be operated only with accessories and spare parts
approved by Sirona. Unapproved accessories and spare parts may lead
to an increased emission of or a reduced immunity to interference.
z
The UNIT should not be operated immediately adjacent to other devices.
If this proves to be unavoidable, the UNIT should be monitored to ensure
that it is used properly.
12.2
Electromagnetic emission
The
UNIT
is intended for operation in the electromagnetic environment
specified below.
The customer or user of the
UNIT
should make sure that it is used in such an
environment.
12.3
Interference immunity
The
UNIT
is intended for operation in the electromagnetic environment
specified below.
NOTE:
XIOS complies with the requirements for electromagnetic compatibility
(EMC) according to IEC 60601-1-2:2001
XIOS is referred to in the following as "UNIT". Observance of the following
information is necessary to ensure safe operation regarding EMC aspects.
Designation of the interface cables
Order No.
PC as peripheral device
USB hub
USB cable, 3m
61 77 401
Emission measurement
Conformity
Electromagnetic environment – guidelines
RF emissions according to
CISPR 11
Group 1
The
UNIT
uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
RF emissions according to
CISPR 11
Class B
The
UNIT
is intended for use in all facilities, including
residential areas and in any facilities connected
directly to a public power supply providing electricity
to buildings used for residential purposes.
Harmonics
according to
IEC 61000-3-2
Class A
Voltage fluctuations/flicker according to
IEC 61000-3-3
Complies