• Swing out the mirror by pressing on
A
.
• Position the patient's head so that the
bite plane is tilt-
ed slightly forward
.
• Switch the
light localizer
on by pressing briefly on the
height adjustment key
↑
.
The light localizer serve for the correct positioning of
the patient.
NOTE
i
Please make sure that the light beam strikes the eyes of
the patient only very briefly.
NOTE
i
The light localizer switches off automatically after about 1
minute.
The horizontal light beam FH
should go through the lower margins of the orbits and the
upper margin of the external auditory orifices (
F
rankfurt
H
orizontal).
• For this reason the FH localizer (9) can be adjusted
manually upwards and downwards.
• Finely adjust the head inclination for the FH line adjust-
ment by tapping the buttons
↑
or
↓
for vertical unit
movement.
The central light beam
should be directed onto the center of the anterior teeth
or the middle of the face.
• Align the center of the anterior teeth or the middle of
the face to the
central light line
.
• Close the temple supports with knob (3).
• Insert contact spacer F (see page 18) and slide down
until it makes contact with the front.
• Swing back the mirror.
• Have the patient take a small step toward the column.
• Recheck the FH position.
A
9
FH
3
F
Summary of Contents for Orthophos 3
Page 2: ......
Page 4: ......
Page 5: ...aÉìíëÅÜ loqelmelp P dÉÄê ìÅÜë åïÉáëìåÖ ...
Page 39: ......
Page 41: ...båÖäáëÜ loqelmelp P léÉê íáåÖ fåëíêìÅíáçåë ...
Page 77: ......
Page 79: ...cê å áë loqelmelp P kçíáÅÉ ÇÛìíáäáë íáçå ...
Page 115: ......
Page 117: ...bëé çä loqelmelp P fåëíêìÅÅáçåÉë ÇÉ ìëç ...
Page 120: ...Sirona Dental Systems GmbH ORTHOPHOS 3 60 82 288 D 3285 4 D 3285 201 04 06 04 ...
Page 153: ......
Page 155: ......