4 Disinfect the mirror sleeve of the CEREC Omnicam with the HLD set
Operating instructions for CEREC Bluecam and CEREC Omnicam Care and HLD (high-level disinfection)
båÖäáëÜ
Remove the mirror sleeve from the CEREC Omnicam
1. Press detent A to release the mirror sleeve.
2. Pull out the mirror sleeve.
3. Place the white protective cap on the camera window and place the
camera in its camera storage unit.
A
Press detent to release
D
Camera window
B
Mirror sleeve
E
Calibration set
C
Sapphire glass (coated)
NOTICE
Risk of damaging the camera window or the coated sapphire glass.
➢ Push the mirror sleeve straight toward the front; do not tilt it.