3 Caring for the CEREC Omnicam and mirror sleeve
Operating instructions for CEREC Bluecam and CEREC Omnicam Care and HLD (high-level disinfection)
båÖäáëÜ
Sterilizing
Sterilizing
The mirror sleeve can be sterilized with hot air (180°C, 30 min) (not in the
autoclave!).
RKI note
Refitting the Mirror Sleeve
Refitting the mirror sleeve
CAUTION
If the CEREC Omnicam accidentally falls down, check to make sure that
the camera window and coated sapphire glass are not damaged. If the
CEREC Omnicam has been damaged, it must no longer be used on
patients.
The CEREC Omnicam must be recalibrated.
NOTICE
Not sterilizable!
Do not under any circumstances sterilize the CEREC Omnicam or the
video cable!
NOTICE
For markets where the RKI* guidelines are to be observed
The mirror sleeve is classified as a "semicritical medical device A"
according to RKI guidelines and therefore does not have to be
autoclavable.
*RKI=Robert Koch Institute, Berlin (Germany).
NOTICE
Risk of damaging the camera window or the coated sapphire glass.
➢ The mirror sleeve must not come into contact with the camera
window.
➢ Push the mirror sleeve straight toward the camera body; do not tilt it.
➢ Carefully refit the mirror sleeve until it locks in place.