9 Dentist element and instrument functions
C3
+
9.7 SPRAYVIT L multifunctional syringe
59 06 404 D 3390
D 3390.201.01.18.02
55
9.7
SPRAYVIT L multifunctional syringe
The multifunctional syringe is used to clean the treat-
ment areas and blow them dry. It supplies preheated air
and water media. The heating cartridges are located in
the handpiece.
IMPORTANT
Also observe the Operating Instructions of the
SPRAYVIT L.
Safety instructions
The SPRAYVIT L is equipped with extensive safety
monitoring functions. However, please observe the
following information.
CAUTION
After changing the SPRAYVIT L hose, no cooling
water for the SPRAYVIT L heating cartridge flows
until the hose is completely filled.
The patient may be scalded by hot steam escaping.
The heating cartridge can overheat and be destroyed.
After changing the hose of the SPRAYVIT L multifunc-
tional syringe, press the Water key
briefly
and repeated-
ly until there is an ample supply of water in the hose be-
fore treating the patient.
CAUTION
If the flow rate in the
SPRAYVIT L
is too low, hot water
may leak out.
The patient can be scalded.
Check the water flow rate before use.
Test the flow rate on a monthly basis and in cases where
you suspect that it may be too low. Clean the nozzle
according to the SPRAYVIT L Operating Instructions.
Instrument light
If the SPRAYVIT L is the only instrument that has been
picked up, you may switch the instrument light on or off.
You can also set the brightness.
The instrument light switches on when you activate the
SPRAYVIT L (if preset).
The instrument light switches off after a time lag of 10 s
when the SPRAYVIT L is no longer activated.
When you place the SPRAYVIT L back in its holder, the
instrument light switches off immediately.
If the SPRAYVIT L is activated together with another
instrument, then the light of the SPRAYVIT L is not
switched on.