2.3
Tango-Utilities
Parameter
Description
Disable Device Calibration
Select this parameter to disable calibration feature on the device.
Disable display setup changes
Select this parameter to lock the display settings.
Disable measurement setup changes Select this parameter to lock measurement recording settings.
Enable fixed record mode
Select this parameter to activate the Record-Mode permanently.
Startup after battery replacement
Automatic start of Tango_Plus after changing the batteries.
Table 2.3: Extended progam settings
2.3.3
Info Tab
You may view the software version of
Tango-Utilities
in the main menu selecting
Help -> About
(figure 2.5).
The firmware version of the device is displayed in the Info-Tab (figure 2.4).
Figure 2.4:
Tango-Utilities
- Info Tab
Figure 2.5:
Tango-Utilities
- About-Box
2.3.4
Setup Tab
The Setup tab of
Tango-Utilities
(figure 2.6) provides the configuration of Tango_Plus for a measurement.
In the column
Display
use the check boxes to enable individual values and comboboxes to set the graphic
refresh rate for display. Several values offer optional parameters, which you may set in the column
Options
.
By the buttons
Open
and
Save
you may open a configuration from the PC or save the current one to it. The
same can be done by using the menu points
File
→
"‘Open Configuration"’
and
File
→
"‘Save Configu-
ration As"’
. The current configuration is transmitted to Tango_Plus by clicking on
Apply
.
In the column
Storage
the correspondig storage conditions are adjusted. The storage can be turned off,
limited to one finish result or done in regular intervals. For the interval time span several values are provided
from 62,5 ms up to 60 min depending on the measurement value. Even if the interval storage is selected a
finish result over the whole measurement time is calculated and stored additionally. This would always be
saved, even if the intervall storage would have been aborted because of full memory. In this case the record
symbol
would stop blinking.
NOTICE!
For the interval storage of the percentiles only the same interval time span
can be selected.
If
Synchronisation
is enabled, the interval storage will be synchronised with the full hours of day time. So
every full hour the current interval is closed and a new one is started. The clock is set every time, when
connecting Tango_Plus to
Tango-Utilities
.
SINUS Messtechnik GmbH
15 of 47