Trouble Shooting.
3.1
LCD Error messages
(CTX-203 and CXT-213 only)
Cadmus devices have an internal self test routine and models fitted with LCD will show any self test failures as below.
Displayed Value
Error Type
Action
Err rtCn
Real Time Clock Error.
Real time clock not set. The clock will continue to function for 15 minutes with
the batteries removed. If batteries are removed or go flat for longer than this
period the clock requires resetting.
Reset clock using Cadmus
Configuration software via the USB
connection.
Err iCAL
Initial Factory calibration error.
Calibration data has been corrupted.
Reset device by placing using
Cadmus-Config or holding the reset
button for 6 seconds. If problem
persists return to supplier for
investigation.
Err rtCE
Real Time Clock Hardware Error.
Real time clock communications error.
Check battery level and replace as
required. If problem persists return to
supplier for investigation.
Err rF
WiFi Hardware Error.
Communications error with WiFi hardware.
Check battery level and replace as
required. If problem persists return to
supplier for investigation.
Err FL U
Serial Flash Hardware Error.
Communications error or unknown device found.
Check battery level and replace as
required. If problem persists return to
supplier for investigation.
Err FL i
Serial Flash Error
Invalid Security ID.
Check battery level and replace as
required. If problem persists return to
supplier for investigation.
Err Fr L
Fram Hardware Error.
Communications error with fram low bank.
Check battery level and replace as
required. If problem persists return to
supplier for investigation.
Err Fr H
Fram Hardware Error.
Communications error with fram high bank.
Check battery level and replace as
required. If problem persists return to
supplier for investigation.
Err USb
USB Hardware Error.
Return to supplier for investigation.
Err FACt
Factory Configuration Error.
Check battery level and replace as
required. If problem persists return to
supplier for investigation.