![Sigma 146774 Operating Manual Download Page 52](http://html1.mh-extra.com/html/sigma/146774/146774_operating-manual_1267883052.webp)
Refrigerated Centrifuge with Heating System Sigma 3-30KHS
6 Using the centrifuge
52 / 91
Version 09/2014, Rev. 1.15 of 08/05/2020
• sb
Translation of the original operating manual, part no. 0701402
Pos : 145 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/060 Betrieb_Spi nc ontr ol L+S/ Spinc ontrol S (Z ent7)/ 060- 0030-0020-0100 Kontr ast änderung @ 27\m od_1405320232376_68.docx @ 193431 @ 4 @ 1
→
6.3.2.11 Changing the contrast
To change the contrast:
• Press and hold the stop button and turn the function knob one notch to
the left. A dialog box will be displayed once the stop button is released.
• Adjust the contrast of the centrifuge display and confirm the change.
Fig. 24: Dialog box for changing the contrast
Pos : 146 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 147 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/060 Betrieb_Spi nc ontr ol L+S/ Spinc ontrol S (Z ent7)/ 060- 0030-0030 Programmbetri eb @ 27\m od_1405320235093_68. docx @ 193459 @ 3 @ 1
→
6.3.3
Program mode
A program contains all of the data that are required for a centrifuge run.
Certain sedimentation results can be repeated under identical conditions.
Programs can be loaded, executed, edited, and deleted when the
centrifuge is at a standstill.
A maximum of 60 programs can be stored under the numbers 1-60. The
program "RAPID_TEMP" does not occupy any storage location and cannot
be deleted. It is used to bring the centrifuge to a specific temperature
without any vessels.
"--" means that the values that are currently set are not a stored program.
The programs can be protected against unauthorised use, modification, or
deletion with the aid of a code (see chapter 6.3.2.7 - "Parameters menu").
Stored programs are listed in the "Process library" menu
.
Pos : 148 /010 Univ ers almodul e/ Seit enwechs el @ 0\mod_1202116244312_0.docx @ 105 @ @ 1