6 Using the centrifuge
32 / 91
Version 09/2014, Rev. 1.15 of 08/05/2020
• sb
Translation of the original operating manual, part no. 0701402
Centrifugation with low capacity
• Install the sample vessels in a rotationally symmetrical manner so that
the buckets and their suspensions are loaded evenly.
• It is not permissible to load angle rotors on only one axis.
correct
incorrect
Fig. 8: Permissible and impermissible loading of an angle rotor and a swing-out rotor
(example illustrations)
NOTE
Pay attention to the marking of the centrifuge (see the illustration below)!
Safety indications on the centrifuge must be kept readable at all times. If
necessary, they must be replaced.
Fig. 9: Safety indication on the centrifuge: Loading of a swing-out rotor