6.2.2
Installation of rotors and accessories
Pos : 109 /100 Sigma/110 BA Z entrif ugen Sigma (Proj ekt e)/ 3-30KS_3- 30KHS/060 Betri eb/060-0020-0020- 0010 Ei ns etz en ei nes St andar drotors 3- 30KS_KH S @ 29\mod_1405321908744_68. doc x @ 198023 @ 4 @ 1
→
6.2.2.1
Installation of a standard rotor
• Open the centrifuge lid by pressing the lid key.
• Unscrew the rotor tie-down screw from the motor shaft (counter-
clockwise).
• Lower the rotor with its central bore straight down onto the motor shaft.
• Tighten the rotor tie-down screw clockwise with the supplied rotor
wrench with 7.5 Nm. In doing so, hold the rotor at its outer rim.
• Follow the safety instructions and hazard warnings (see chapter 3 -
WARNING
Once a day or after 20 cycles, the rotor tie-down screw must be loosened
by some turns, and the rotor must be lifted and fastened again. This
ensures a proper connection between the rotor and the motor shaft.
CAUTION
When using rotors for microtiter plate formats:
Ensure that the plate holders are inserted together with the plates into the
buckets.
NOTE
The lid screw serves for the fastening of the lid onto the rotor only, not for
the fastening of the rotor onto the motor shaft.
Removal of a rotor
• Loosen the rotor tie-down screw by turning it anti-clockwise and remove
the rotor.