6.2.2.2
Installation of a standard angle rotor with a hermetic lid
• Open the centrifuge lid by pressing the lid key.
• Unscrew the rotor tie-down screw from the motor shaft (counter-
clockwise).
• Screw the rotor lid onto the rotor and tighten it.
• Lower the rotor with the lid with its central bore straight down onto the
motor shaft.
• Insert the rotor tie-down screw into the motor shaft. Tighten the rotor tie-
down screw clockwise with the supplied rotor wrench with 7.5 Nm. In
doing so, hold the rotor at its outer rim.
• Follow the safety instructions and hazard warnings (see chapter 3 -
NOTE
The rotor can also be used without a cover.
NOTE
The lid screw serves for the fastening of the lid onto the rotor only, not for
the fastening of the rotor onto the motor shaft.
Removal of a rotor
• Unscrew the rotor tie-down screw (counter-clockwise) and remove the
rotor with the lid.
• Open the lid by turning it counter-clockwise and remove the lid.
1
Rotor tie down screw
2
Lid
3
Rotor
Fig. 6: Angle rotor with a hermetic lid
Pos : 112 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 113 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/060 Betrieb_Spi nc ontr ol L+S/ 060- 0020-0020-0030 Eins et zen eines R ot ors mit g-lock-Sy st em @ 7\m od_1306402032880_68.docx @ 43576 @ 4 @ 1
→