SSG5000A RF Signal Generator User Manua
l
SSG5000A User Manual 46
⚫
Train: Multiple pulse signals in one pulse period. At this time, the pulse sequence
setting item appears, and the two settings “Double Pulse Delay” and “#2 Width” are
hidden.
Note:
This setting item will be hidden when the modulation source is
“Ext”.
Pulse Period
The pulse period represents the time interval between two adjacent periodic pulses. The
period of the single pulse or double pulse modulation signal can be set by touching the
screen keyboard or pressing the front panel keypad.
Note:
This setting item will be hidden when the modulation source is “Ext” or the pulse
mode is
“
Train
”
.
Pulse Width
The pulse width indicates the duration when the pulse signal is at a high level. The width
of the first pulse of the single pulse modulation signal or the double pulse modulation
signal can be set by touching screen keyboard or pressing the front panel keypad.
Note:
This setting item will be hidden when the modulation source is “Ext” or the pulse
mode is “Train”.