SSG5000A RF Signal Generator User Manua
l
19 SSG5000A User Manual
Note:
⚫
The relationship between the RF output frequency, the display frequency and
frequency offset is:
Display frequency = RF output Fre Frequency offset
⚫
The difference value between the display frequency and the frequency offset is the
actual RF output frequency, which must not exceed the instrument frequency output
range.
Phase Offset
Setting the phase offset from the RF signal
Application: When multiple RF sources are outputting RF signals at the same time,
multiple signals can be output to the same phase or a fixed phase offset by adjusting this
parameter.
Also you should set the devices the same frequency and synchronous them.
1. Keypad operation
Press
FREQ
->
Phase Offset
to make the focus fall on this parameter item.
⚫
Use the numeric keypad to input the desired value and press the unit button to select
the desired unit. Optional units are deg and rad. Press
ENTER
to select the current
unit.
⚫
Press
ENTER
or press the knob to enter the parameter editing state, and then use
the left and right direction keys to move the cursor to the desired position. Press the
up and down direction keys, rotate the knob or press the numeric keypad to modify
the value. Press
ENTER
, knob or
ESC
to exit the editing mode.
2. Touch screen operation
Click on
RF
->
FREQ
->
Phase Offset
to enter the interface of phase offset setting.
⚫
Click the phase offset parameter to pop up the keypad interface, and then click the
number symbol in the keypad to change the value. Click the unit to modify the unit.