SDS2000X Plus Series Digital Oscilloscope User Manual
139 / 335
WWW.SIGLENT.COM
If you set the trigger condition to 7 address & data
or 10 address & data:
Address
can be selected in the hexadecimal
range of 0x00 to 0x7F (7-bit) or 0x3FF (10-bit).
If the address is selected as "0xXX (7-bit
address)" or "0xXXX (10-bit address)", the
address is ignored. It will always trigger on
the Ack bit followed by the address.
Data1
and
Data2
can be selected in the
hexadecimal range of 0x00 to 0xFF. If the data
is selected as "0xXX", the data is ignored. It
will always trigger on the Ack bit followed by
the address.
R/W Bit
can be specified as Write, Read or
Don't Care.
Data Length — The data length range is 1 to 12 bytes. When the SDA data
length is equal to the value of the Data Length setting and the address
length is the same as the set value, the oscilloscope will be triggered.
Touch the
Address Length
to select "7-bits" or "10-bits" to match the
address of the input signal.
Touch the
Data Length
and turn the universal knob or by virtual
keypad to set the data length to match the data length of the input
signal.
Summary of Contents for SDS2000X Plus Series
Page 1: ...SDS2000X Plus Series Digital Oscilloscope User Manual UM0102XP E01B ...
Page 225: ...SDS2000X Plus Series Digital Oscilloscope User Manual 225 335 WWW SIGLENT COM editor ...
Page 274: ...SDS2000X Plus Series Digital Oscilloscope User Manual 274 335 WWW SIGLENT COM ...
Page 292: ...SDS2000X Plus Series Digital Oscilloscope User Manual 292 335 WWW SIGLENT COM M2 Mode M3 Mode ...
Page 335: ...SDS2000X Plus Series Digital Oscilloscope User Manual 335 335 WWW SIGLENT COM ...