SDM3055 Digital Multimeter 29
To Measure DC Voltage
Range:
200mV, 2V, 20V, 200V, 1000V
Max Resolution:
100nV (in the range of 200mV)
Input Protection:
a 1000 V protection is available in all ranges and a 10%
overrange for all ranges except 1000 V range. If the reading exceeds the
range,
“
overload
”
will be displayed.
Operating Steps:
1. Enable the DCV measurement
Press
on the front panel to enter the DC Voltage measurement
interface, as shown in Diagram 2- 9.
Diagram 2- 9 DC Voltage Measurement Interface
2. Make connection
Connect the test leads with the measured signal by referring to
“Measurement Connections”.
3. Set the range
Press
【
Range
】
to select a range for the measurement. You can also use
the
,
, and
keys on the front panel to select the range.
Auto (autorange) automatically selects the range for the measurement
based on the input. Autoranging is convenient, but it results in slower
measurements than using a manual range. Autoranging goes up a range
at 110% of the present range, and down a range below 10% of the
present range.
4. Set the intergration