![SIGLENT SDM3065X User Manual Download Page 37](http://html.mh-extra.com/html/siglent/sdm3065x/sdm3065x_user-manual_1267453037.webp)
SIG
L
ENT
SDM3055 Digital Multimeter 25
AC Filter
AC filter applies to ACV and ACI measurements. It can optimize the
low-frequency accuracy and minimize the AC settling time. SDM3065X
provides three types of AC filters (>3Hz, >20Hz, >200Hz).
The AC filter to be used is determined by the input signal frequency. You
should generally select the highest frequency filter whose frequency is less
than that of the signal you are measuring, because higher frequency filters
result in faster measurements. For example, when measuring a signal
between 20 and 200 Hz, use the 20 Hz filter. If measurement speed is not an
issue, choosing a lower frequency filter may result in quieter measurements,
depending on the signal that you are measuring.
Press
【
Filter
】
in the menu of ACV or ACI measurement to show the setting
options, as shown in Diagram 2-6 (take ACV measurement for instance).
Then press the menu operation key to activate the corresponding
configuration.
Diagram 2-6 AC Filter setting interface