SDM3055 Digital Multimeter 21
Intergration Time and Resolution
Integration time is the period during which the multimeter‟s analog-to-digital
(A/D) converter samples the input signal for a measurement. The longer the
integration time is, the slower the measurement speed will be and the higher
the resolution will be; the shorter the integration time is, the faster the
measurement will be and the lower the resolution will be. The integration time
applies to DCV, DCI, 2WR and 4WR measurements.
SDM3065X express the integration time by the power line cycles, the unit is
PLC. The multimeter automatically detects the input power frequency at
power-on. I f the frequency is 50Hz, the intergration time can be set to
0.005PLC, 0.05 PLC, 0.5 PLC, 1 PLC, 10 PLC, 100 PLC and the default is 10
PLC. If the frequency is 60Hz, the intergration time can be set to 0.006PLC,
0.06 PLC, 0.6 PLC, 1 PLC, 10 PLC, 100 PLC and the default is 10 PLC .
SDM3065X holds reading resolutions of 4½ , 5½ and 6½ digits. It automatically
selects a reading resolution according to the current measurement settings.
1.
In DCV, DCI and OHM measurements, press
【
Aperture
】
to set the
intergration time, as shown in Diagram 2-5 (take DCV measurement for
instance). The integration time affects the resolution.
Diagram 2- 3 Intergration time selection menu