Menu diagrams
A.10 External input
80
A.10
External input
1)
The value showing totalizer 1 on the display is frozen for as long as the digital output is
activated. However, totalizer 1 continues counting, and when the digital input is released, the
value on the display again follows totalizer 1.
2)
MAG 6000 I cannot be equipped with cleaning unit. The cleaning option for relay output is
however possible. When selecting function for MAG 6000 I relay output, the relay output has
the same behavior as if cleaning unit was installed.
The relay output of the transmitter determines when the relay is on by applying voltage for
approximately 60 seconds. The metering is resumed after another 60 seconds when the cycle
is complete. (The display is locked during this time). The time cycle can be set at 1 to 240
hours. If the cycle is set at for example three hours, the transmitter will be active every three
hours.