94
11.7.3
Procedure when Using the SIMATIC FS600 Parameterization Software LS4soft
User Software Version 1.09 or later
1) Starting the PCs
2) Connection of X1 (electrical power supply with zone pair activated) and
X2 ( RS232 cable 1:1 )
3) Start SIMATIC FS600 Parameterization Software LS4soft user software
4) Select the “Authorized Customer” authorization level
5) Confirm the echo data shown by the scanner
6) Click on the “Calibrate optical window monitoring” icon under “System data”
7) Start the optical window calibration with “Calibration” (lasts a few seconds, typically
with values between 100 - 700).
8) Click on the “Close” button