Installation
3 - 5
Siemens AG
SPR2-240.841.01
Page 5 of 10
POWERMOBIL
Medical Solutions
Rev. 01
05.03
CS PS 24
Threading in C-arm
3
•
To facilitate threading in the C-arm unscrew the holders for the covers of the cable mod-
ule present on both sides (1/Fig. 12).
•
The cable module is still blocked (Fig. 1).
•
There are 2 rubber buffers attached to the cable module (1/Fig. 13). If the new cable
module does not include these buffers, remove them from the old cable module and at-
tach them to the new module.
•
Lift the C-arm with 2 persons, and turn it into the bearing support (Fig. 14).
To avoid high pressure on the bearing rollers and the C-arm make sure that the C-arm is
not inserted canted.
•
Refit the holders on both sides (1/Fig. 12).
Fig. 12
Fig. 13
Fig. 14