
POWERMOBIL
SPR2-240.841.01
Page 4 of 12
Siemens AG
Rev. 01
05.03
CS PS 24
Medical Solutions
2 - 4
Disassembly
POWERPHOS
2
•
If applicable, dismantle the DIAMENTOR dose measuring chamber and loosen the con-
nector. Also remove the insulating disk underneath the DIAMENTOR dose measuring
chamber.
•
Remove the cover above the collimator. The collimator remains attached to the
POWERPHOS.
•
Remove the fan component from the POWERPHOS.
•
Remove the cable ties of the fan cable.
•
Note down the connection order of the fan assembly supply cable, and loosen the ca-
bles (terminals and colors of the cables, see 1/Fig. 6).
•
Also loosen the cable for the fan integrated in the C-arm. Also note down the connection
order (1/Fig. 6).
Fig. 6
Fig. 7
Fig. 8