Disassembly
2 - 9
Siemens AG
SPR2-240.841.01
Page 9 of 12
POWERMOBIL
Medical Solutions
Rev. 01
05.03
CS PS 24
Cable guide in the horizontal carriage
2
•
Remove the cover above the cable (1/Fig. 17). Also loosen the cable clamps (2/Fig. 17).
•
Unscrew the two cable guides, see (1/Fig. 18) and (2/Fig. 18).
•
Unscrew the cable guide (1/Fig. 19).
•
Unscrew the cable guide (1/Fig. 20). After moving the horizontal carriage, the second
screw of the cable guide will be accessible.
Fig. 17
Fig. 18
Fig. 19