
Disassembly
2 - 3
Siemens AG
SPR2-240.841.01
Page 3 of 12
POWERMOBIL
Medical Solutions
Rev. 01
05.03
CS PS 24
X-ray image intensifier and I.I. mini voltage supply
2
•
Move the X-ray image intensifier into service position. (See chapter 1 and Fig. 4 in the
service manual.)
•
Disconnect the I.I. mini voltage supply (high-voltage section and control section).
(See chapter 6 in the service manual. Please observe the safety information on dis-
charging the I.I. mini voltage supply.)
•
Disconnect the HV cable, as well as E3, E2 and E1 cables.
•
Disconnect the CON1 connector.
•
Eject the black cathode cable of the X-ray image intensifier (CON1.8)
•
Remove the protective ground wire from the X-ray image intensifier and disconnect the
I.I. mini voltage supply.
•
Disconnect the M4 X1 camera plug from the support bracket, and remove it.
•
Disconnect the M30 X1 plug.
•
Loosen the cable ties. All image intensifier cables are now disconnected.
•
Remove the X-ray image intensifier from the C-arm, and store it in a dark place
(avoid exposure to direct sunlight). Avoid exposing the CCD sensor to light and dirt by
covering the camera PC board.
•
Remove the I.I. mini voltage supply from the C-arm.
•
The X-ray image intensifier section of the C-arm contains a round LEMOSA plug con-
nection. Loosen this pug connection (Fig. 5).
Fig. 4
Fig. 5