24
General Information
POWERMOBIL
SPR2-240.831.01.02.02
Siemens AG
07.05
CS PS 24
Page 24 of 44
Medical Solutions
3.2.2 Equivalent leakage current measure-
ment
n.a.
3.3 Patient leakage current or equivalent
patient leakage current measurement
n.a.
3.3.1 Patient leakage current measurement n.a.
3.3.2 Equivalent patient leakage current
measurement
n.a.
4. Mechanical checks
n.a.
4.1 Are all wall, ceiling and floor mountings
secure and undamaged?
3. Mechanical safety, check:
SIM Monitor(s) / displays
4.2 Are all mechanically moved system
parts clean and running smoothly (lubri-
cated, if required)?
3. Mechanical safety, checks:
SIM Covers
SIM C-arm
SIM Lifting column
4.3 Are the wire cables, belts and spindles
free of wear and tear?
3. Mechanical safety, checks:
SIM Lifting column
4.4 Mobile equipment: Are the wheels, roll-
ers and brakes OK?
3. Mechanical safety, checks:
SIM Foot brake
SIM Brakes SIM Wheels
4.5 Is there any unusual noise during oper-
ation (e.g., gearing)?
3. Mechanical safety, checks:
SIM C-arm
SIM Lifting column
5. Functional checks
n.a.
5.1 Function of the Emergency Stop switch 3. Mechanical safety, check:
SIM Emergency Stop switch
5.2 Are the warning devices functioning
properly?
3. Electrical safety, checks:
SIE Audible warning signal
SIE Radiation indicators
5.3. Do all system movements stop prop-
erly in their end positions?
3. Mechanical safety, check:
SIM Lifting column
5.4 Does the collision protection device
stop all system movements properly (e.g.
collision protection)?
Check the collision protection, if it is
installed.
Otherwise: n.a.
Technical safety checks, test certificate
of the check
Maintenance instructions for the
check/comments