Siemens AG
SPR2-240.831.01.02.02
POWERMOBIL
07.05
CS PS 24
General Information
23
Page 23 of 44
Medical Solutions
2.3 Are undamaged accessories being
used?
3. Mechanical safety, checks:
SIM I.I. laser light localizer mechanics
SIM I.I. laser light localizer function
SIM DICOM bridge mounting
SIM Codonics printer mounting
2.4 System radiation protection devices
present and undamaged? (No on-site radi-
ation shield)
3. Electrical safety, checks:
SIE Fluoroscopic timer
SIE Audible warning signal
SIE Compulsory radiation switch-off
SIE Check of the radiation release
switches
SIE Dose rate
SIE Radiation indicators
SIE Iris collimator
SIE Area dose product measuring unit
2.5 Are the required operator documents
complete, present and legible?
Operating instructions must be available.
Ask the customer for additional required
operator documents and check them for
completeness, availability, and legibility.
2.6 Are all warning labels in place and rec-
ognizable?
3. Mechanical safety, checks:
SIM Warning signs
SIM ID labels
2.7 Are the operating symbols, light indica-
tors and button labels OK?
3. Electrical safety, check:
SIE Check of the operating functions
3. Electrical checks
n.a.
3.1 Measurement of protective conductor
resistance
3. Electrical safety, check:
SIE Protective conductor test
5.1 Final work steps, check:
SIE Protective conductor test
3.2 Leakage current or equivalent leakage
current measurement
n.a.
3.2.1 Leakage current measurement
3. Electrical safety, check:
SIE Leakage current
5.1 Final work steps, check:
SIE Leakage current
Technical safety checks, test certificate
of the check
Maintenance instructions for the
check/comments