
5 - 1
Siemens AG
RXB8-115.061.02
Page 1 of 6
POLYMOBIL III
Medical Solutions
Rev. 01
10.95
CS PS 24
Replacing important components
5
Replacing the single tank
5
To replace the single tank, remove it from the mobile stand as follows:
•
Move the arm system in the lowest position and check if the safety locking (pin 1/Fig. 1)
has engaged.
The stand cannot be moved upwards any longer.
•
Tighten the turning knob(3/Fig. 1).
•
Loosen the threaded ring and pull out the connector of the single tank.
•
Using an 8 mm Allen key, loosen the two screws on the bracket (2/Fig. 1) approx. two
turns.
•
Lift the single tank off vertically and place it on a soft base.
•
Detach the collimator (see chapter 5, page 2).
•
Install the new single tank in the reverse order.
•
Check alignment of light field to radiation field and adjust if necessary. (see "Aligning the
light field to the radiation field").
Fig. 1
1
2
3