POLYMOBIL III
RXB8-115.061.02
Page 18 of 24
Siemens AG
Rev. 01
10.95
CS PS 24
Medical Solutions
4 - 18
Notes on trouble-shooting
Checking the mAs values
4
•
Remove the 4 Allen screws ( 3 mm key size) from the lid of the single tank and take off the
lid (Fig. 2).
•
Remove jumper ”mA+ / mA - ” in the single tank.
•
Connect the mAs meter to the ”mA + / mA - ” sockets (X038I-5/5).
•
Trigger the following exposures:
•
Remove the mAs meter and reinsert the jumper in the single tank at ”mA+ / mA - ” .
•
Reinstall the lid of the single tank.
Setting at control panel
valid mAs values
40 kV, 200 mAs
195 ... 205 mAs
66 kV, 100 mAs
97 ... 103 mAs
90 kV, 1 mAs
1,0 ... 1,3 mAs
Fig. 2