Notes on trouble-shooting
4 - 9
Siemens AG
RXB8-115.061.02
Page 9 of 24
POLYMOBIL III
Medical Solutions
Rev. 01
10.95
CS PS 24
Checking the maximum main inverter frequency
4
•
POLYMOBIL
OFF
•
Open bridge "S1" on D920 (solder).
•
Connect oscilloscope on D910 to MP "REG" and "GND".
•
POLYMOBIL
ON
•
Trigger exposure.
F
max
= 10 kHz
T
min
= 100
µ
s Tolerance: + 0
µ
s / - 5
µ
s
•
Adjust the maximum main inverter frequency with potentiometer P2 to D910
(X028I-91E).
•
POLYMOBIL
OFF
•
Connect jumper "S1" on D920 (solder).
- Err 15 is displayed.
- No radiation.
NOTICE
100