Notes on trouble-shooting
4 - 21
Siemens AG
RXB8-115.061.02
Page 21 of 24
POLYMOBIL III
Medical Solutions
Rev. 01
10.95
CS PS 24
Fig. 4
Evaluation:
•
Measure the deviations between light field edges and radiation field edges on all four
sides (X1,X2,Y1, Y2) as shown in Fig. 4.
•
Determine the total deviations in the X and Y direction (ignore the +/- signs).
•
Both the length deviation (
Σ
Y) and the width deviation (
Σ
X) must be less than 1.6 cm.
X
1
Y
1
Y
2
X
2
Film
Light field
Radiation field