
4 - 1
Siemens AG
RXB8-115.061.02
Page 1 of 24
POLYMOBIL III
Medical Solutions
Rev. 01
10.95
CS PS 24
Notes on trouble-shooting
4
In case of errors which impair radiography, perform the following checks.
Checking the line voltage, fuses and LEDs
4
a) Checking the line voltage
•
Measure the supply voltage on site at the safety socket using the digital multimeter.
•
Plug the power plug of the POLYMOBIL into the safety socket.
•
POLYMOBIL
ON
•
POLYMOBIL
OFF
.
For voltages > 240 V AC see Startup Instructions
RXB8-115.034.02... chapter 3.
If the POLYMOBIL cannot be switched on although the power plug
and power voltage have been connected, the overcurrent circuit
breaker F1 may have responded (automatic circuit breaker
1/Fig. 1)
NOTICE
NOTICE