Service Instructions
Multimobil 5C
Siemens Ltd. Med India
Version 5.0
Page 51 of 73
Copyright
©
SIEMENS LTD. All rights reserved. For internal use only
8.1.2 No X-ray
START
IS UNIT IN
STANDBY STATE?
INITIALISATION ERRORS
IS TEMP INDICATOR ON
THE TOP PANEL ON?
WAIT FOR SINGLE TANK
COOLING TIME
HAS THERMAL SWITCH
FUNCTIONED AT > 65 ‘C
GO TO RADIOGRAPHY MODE
SELECT HAND HELD RAD
SWITCH
GO TO FLUORO MODE & SELECT FOOT
SWITCH AND OR HAND HELD FLUORO
SWITCH
IS FILAMENT
BOOSTED & KV
GENERATED?
FLUOROSCOPY
ERRORS
IS X-RAY NOT
PRESENT IN FLUORO
RADIOGRAPHIC
ERRORS
IS ERROR
DISPLAYED?
CONFIRM
SWITCH
FUNCTIONING UPTO
MICROCONTROLLER
No
MEASURE DOSE AND
CONFIRM X-RAY
GENERATED.
A
Yes
Yes
No
Yes
No
No
Yes
No
Yes
No
Yes
THERMAL SWITCH
FUNCTIONS UPTO
MICROCONTROLLER
Yes
No
REPLACE SINGLE TANK OR
D508 PCB OR D915 PCB