Service Instructions
Multimobil 5C
Siemens Ltd. Med India
Version 5.0
Page 9 of 73
Copyright
©
SIEMENS LTD. All rights reserved. For internal use only
2 Servicing
2.1 Tools and measuring instruments required
•
Usual service tools
•
Digital Multimeter
•
2 channel storage oscilloscope
•
mAs meter
•
Megger
•
Radiation detector
•
Mechanical tool kit containing set of spanners, screwdrivers, Allen keys etc.
During Oscilloscope operation the protective ground
wire connection in the power cable must not be
interrupted under any circumstances.
2.2 Replacing damaged or missing screws
Damaged or missing screws should only be replaced with screws having similar
parameters as that of the original. An ornamental screw should be replaced with an
ornamental screw only for retaining the aesthetic value where as a nickel-plated
screw/bolt should be replaced with a nickel plated screw/bolt with same specification
(size, material and tensile strength) as that of the original screw/bolt.
2.3 Cleaning and Disinfecting
The unit must always be switched OFF or disconnected before cleaning.
Use only water or a Luke warm mixture of a household cleaner diluted with water to
clean the unit.
Do not use an abrasive or organic solvents or cleaning agents containing solvents
such as gasoline used for cleaning purposes, alcohol or stain remover. Do not spray
water on the unit.
2.3.1 Cleaning the Unit
Switch OFF the system prior to cleaning, disinfecting, or sterilizing.
Clean the stand with a damp cloth or cotton pad.
Dampen the cloth or pad with water or lukewarm, diluted commercially available liquid
cleaning solution.
Do not use scouring powder, organic solvents (may
damage materials) or solvent based cleaning fluids
(benzene, alcohol, spot remove).
Do not spray on the unit. The cleaning fluid must not seep
into the generator.
WARNING
CAUTION