Service Instructions
Multimobil 5C
Siemens Ltd. Med India
Version 5.0
Page 21 of 73
Copyright
©
SIEMENS LTD. All rights reserved. For internal use only
5.1.2 Checking the fuses
Open the front cover of the control unit.
Loosen the Fuse carrier, which are mounted on the base of the control unit and check
the continuity of fuse link. If fuse link has responded for over current replace the Fuse
Link.
5.1.2.1
Fuses on Transformer Plate
Identification Rating
Purpose
U1
16 A Delayed
Mains Fuse
U2
16 A Delayed
Mains Fuse
U3
6A Delayed, Glass cartridge
Motor Supply
U4
6A Delayed, Glass cartridge
Motor Supply
U5
6A Delayed, Glass cartridge
Motor Supply
U6
6A Delayed, Glass cartridge
Charging
resistor
to
capacitor bank
U7
6A Delayed, Glass cartridge
SMPS supply SMPS 1 & 2
U8
6A Delayed, Glass cartridge
SMPS supply D982 & D985
Note: All Fuses are rated for 250V
5.1.2.2
Fuses on D506
Identification Rating
Purpose
F1
250 mA Delayed, Glass Cartridge T1 transformer primary
F2
630 mA Delayed, Glass Cartridge 240/264 V AC Supply
F3
630 mA Delayed, Glass Cartridge 24 V for K3, K4, K5 & K6
relays
F5
4 A Delayed, Glass Cartridge
Filament Circuit
Note: All Fuses are rated for 250V
5.1.2.3
Fuses on D508
Identification Rating
Purpose
F2
100 mA Delayed, Glass Cartridge Collimator power supply
Note: All Fuses are rated for 250V