6.2 Checks
SIE
Emergency STOP
•
Press Emergency STOP, see chapter "Compression and system movements"
.
- Are compression, rotation and vertical travel of lifting carriage blocked?
•
Release Emergency STOP.
PMF
Indicators
•
Check the LEDs on the operating console by starting the Panel Test via the service PC.
PMP
Error memory
•
Read out the error memory and the exposure counter with the service PC.
PMP
Record error memory
PMP
Delete the error memory
•
Enter the data of the exposure counter and the error memory in the protocol and delete
the error memory o n l y.
PMF
Auxiliary voltages
•
With the voltmeter, measure the auxiliary voltages according to circuit diagram
(SPB7-250.844...).
•
System OFF
PMF
OPDOSE
•
Check the selection of various anodes / filter combinations and OPDOSE according to
the Installation Start-Up instructions, chapter "Testing and adjusting the OPDOSE".
PMA
UIs and SPEED Infos
•
Check whether all relevant UIs and SPEED Infos have been completed.