Compression and system movements
4.1
Testing the “compression” function
•
Attach the spring scale as shown in
and make sure not to damage the
compression plate during the following tests:
Tolerance:
±
1 kp.
4.1.1 Compression
cutoff
•
System ON
PMF
Presetting
•
Select any value (e.g. 6 kp) on the potentiometer
and compress. Does
the compression unit cut off at that setting?
SIE
Max. value for cutoff
•
Select the maximum value (20 kp) with the potentiometer
. (20 kp).
Does the unit cut off when it reaches the maximum value?
•
Correct display (chapter 2 "System",
).
PMF
OPCOMP
•
Position your forearm between the object table and the compression plate and check
the OPCOMP function.
¹
Typically, the cutoff value is between 6 and 10 kp with the factory setting.
Fig. 1:
Compression