Siemens AG
SPB7-250.831.01.02.02
MAMMOMAT Novation
08.04
CS SD 24
System
13
Page 13 of 32
Medical Solutions
2-
2
System
2.1
Checks
PMP
Radiation protection
•
Check the radiation protection shield for possible damage
.
•
Information regarding the lead equivalent value must be legible.
PMP
Face shield and compression plate
•
Check the face shield and compression plate for correct seating and for damage
.
SIM
Level
•
Verify that the system is leveled with respect to the floor. Check the leveling screws
with a water level and, if available, the floor mounting (optional).
SIM
Swivel arm attachment
•
Check the swivel arm attachment by pulling on both support rails (minimum play of ap-
proximately 2° is correct).
.
SIM
Basic table
•
Check the mounting of the basic table as well as the locking mechanism for the expo-
sure system
.
Fig. 1:
MAMMOMAT Checks