Test exposures
MAMMOMAT Novation
SPB7-250.831.01.02.02
Siemens AG
08.04
CS SD 24
Page 24 of 32
Medical Solutions
5.1
Evaluating test values
PMF
Anode acceleration
•
Listen to the sound generated by the accelerating anode. Is the sound normal?
PMF
kV and mA
•
Do the measured kV and mA values agree with the values selected?
PMF
mAs selection
•
Do the measured mAs values agree with the values selected?
PMF
mAs value displayed
•
Does the mAs value measured agree with the mAs value displayed (for AEC)?
PMF
Grid voltage
•
Attach the magnification table. Measure the grid voltage (small focus) on tube test point
G against test point H
3
. Does the value measured agree with the value in the test proto-
col?
•
Test both focal spots (if available).
PMF
Signal lamp
•
Does the radiation-ON indicator light up?
PMF
Chest wall missed tissue
•
Perform the check according to the Quality Control Manual, chapter "Chest wall missed
tissue”, to ensure that the chest wall missed tissue is within the tolerance of max. 5 mm.
X
PMF
Spatial resolution
•
Perform the check according to the Quality Control Manual, chapter "Spatial resolu-
tion", to ensure that a sufficient spatial resolution is obtained with the system.
X
PMF
Mean glandular dose
•
Perform the check according to the Quality Control Manual, chapter "Mean glandular
dose, to ensure that the mean glandular doses are according to regulations.
X