16
Column stand
MAMMOMAT Novation
SPB7-250.831.01.02.02
Siemens AG
08.04
CS SD 24
Page 16 of 32
Medical Solutions
3.1 Checks
SIM
Steel ropes
•
Do the steel ropes (balancing and flap), the drive belt, the suspension or the pulleys
show any damage (e.g. fraying or splitting) or signs of wear?
PMA
Oil and grease
•
If they are okay, lightly oil the wire cable and grease the rails, then remove the protec-
tive strips.
SIE
Limit switches
•
System ON. Move the lifting carriage up and down several times, the lifting carriage
should be in an upright (0°) position.
NOTE
It is normal to hear some noise caused by the rotation of the radi-
ation filter in the collimator when switching the system ON.
•
Do the limit switches stop the carriage travel at the top and at the bottom?
•
Activate safety switch S882. System movements must be blocked.
PMF
Vertical travel
•
Is the vertical travel smooth and noiseless?
•
Switch the system OFF and attach the protective strips once again.
SIM
Safety catch and rotation safety catch
•
Do the safety catch
and rotation safety catch
show any
visible damage? Are the springs in good condition?
Fig. 4:
Catch
SIM
Ball bearings
•
The lifting carriage is equipped with ten or twelve ball bearings (refer to
).
Check that all ball bearings are in place and that they show no signs of damage.