22
Troubleshooting
Error Code
Severity
Possible Cause
Corrective Action
ASL - Autoslide is not available to make slides
Warning
Autoslide is not online and ready for use or ASL is
not enabled
Check Status Screen to determine if daily or weekly startup
completed and troubleshoot as necessary
ASL - Please Close manual door
Warning
Stat door is open
Close the manual inlet / stat door
Autoslide Command Error
Warning
Autoslide command error
Retry the command
Autoslide Daily Shutdown Incomplete
Warning
An error occurred that prevented the autoslide daily
shutdown from completing
Check message log for errors, correct any problems
Perform the daily shutdown. If still not successful, perform a
reset and perform the daily shutdown
Autoslide Daily Startup Incomplete
Warning
An error occurred that prevented the autoslide daily
startup from completing
Check message log for errors, correct any problem
Perform daily startup. If still not successful, perform a reset
followed by the daily startup
Autoslide is Processing Slides
Warning
Attempt was made to go to a configuration screen
while slides were being processed
Wait until all slides are completed before attempting to
change configuration
Autoslide Manual Inlet Door
Cannot be Opened
Warning
The manual inlet door cannot be opened because
the stainer is not enabled for use or the ADVIA 2120
Autosampler is active
Enable the stainer in the Autoslide options menu or stop
Autosampler if appropriate, and try again
Note:
If the Autoslide is powered off, please wait 30
seconds before the power is switched back on.
This allows the ADVIA software to timeout
appropriately.