Weekly Shutdown - Startup
1. Select
Procedures > Clean Autoslide >
Weekly Shutdown
2. Follow the on screen instructions. Remove
stain and buffer lines from reagent bottles
and place in empty containers. Use separate
containers for buffer and stain lines.
3. Select
Next
; the system will perform a 5
minute air prime.
4. When prompted, move buffer and stain lines
to containers containing methanol. Use a
separate container for buffer and stain lines.
5. When complete,
Exit
Weekly Shutdown and
select
Weekly Startup
.
6. Follow the on screen instructions. Place stain
and buffer lines in the empty containers and
select
Next
.
7. When prompted, place lines in the reagent
bottles. Select
Next
.
8. Select
Exit
when Weekly Startup complete.