
13
Empty Glass Particle Tray
1. The Glass Particle Tray is located beneath the
slide dispenser. Using both hands, carefully
pull the Glass Particle Tray forward and
remove tray from Autoslide.
WARNING:
Use caution when handling
objects with sharp edges
2. Discard the glass particles in the tray as per
your laboratory policy.
3. Place the tray back in position on Autoslide
holder.