Page 2
The Probe – INSTRUCTION MANUAL
7ML19985GD61
mmmmm
English
Mounting
Threaded
Flange Adapter (optional)
Sanitary
Note:
Mount The Probe so that the face of the sensor is at least 25 cm above the highest
anticipated level.
The Probe is available in three thread types:
2" NPT, 2" BSP or PF2.
Note:
Before inserting The Probe into its
mounting hole, ensure that the threads are of
the same type to avoid damaging The Probe
threads.
hinged lid
117 mm
(4.6F")
cable entry
’knock out’
87 mm
(3.4") ø
61 mm
(2.4")
7 mm
(0.3")
84 mm
(3.3") max.
279 mm
(11")
electronics
mounting
thread
sensor
53 mm
(2.1")
ø max.
mounting
thread
74 mm
(2.9")
The Probe
optional
flange
adapter
The Probe can be fitted with the optional 75 mm
(3") flange adapter for mating to 3" ANSI, DIN
65PN10 and JIS 10K3B flanges.
hinged lid
117 mm
(4.6")
cable entry
’knock out’
87 mm
(3.4") ø
61 mm
(2.4")
7 mm
(0.3")
84 mm
(3.3") max.
electronics
mounting
thread
sensor
53 mm
(2.1")
ø max.
279 mm
(11")
76 mm
(2.9")
integral
sealing ring
Notes:
• Mount The Probe so that the face of
the sensor is at least 25 cm above
the highest anticipated level.
• The Sanitary Probe 3A approval is
for chemical clean-in-place
applications to 60 °C (140 °F) only.
Ensure your cleaning chemicals are
compatible with PVDF.
Sanitary Ferrule,4" FDA Approved
(US Food and Drug Administration)
119 mm (4.68")
97 mm (3.83")
dimensions are approximate
The Probe
clamp
adjusting
wing nut
ferrule
tank
• mount The Probe onto the top of
the tank’s sanitary ferrule
• secure mating by surrounding
the joint with the clamp
• tighten adjusting wing nut
Note:
Inside of sanitary
ferrule must be smooth, free of
burrs, seams or ridges.
Summary of Contents for 7ml12011ef00
Page 1: ...Instruction Manual February 2008 LEVEL TRANSMITTER the probe English Deutsch Español Français ...
Page 14: ...Page 12 The Probe INSTRUCTION MANUAL 7ML19985GD61 mmmmm English ...
Page 26: ...Seite 12 The Probe BETRIEBSANLEITUNG 7ML19985GD61 mmmmm Deutsch ...
Page 38: ...Página 12 The Probe MANUAL DE INSTRUCCIONES 7ML19985GD61 ...