Technical data
8
8.1
MAG 5100 W
Table 8-1
Technical data
Version
MAG 5100W (7ME6520)
MAG 5100W (7ME6580)
Product characteristic
Mainly for the European market
EPDM or NBR lining
Mainly for the non-European market
Ebonite lining
Design and nominal size
Coned sensor:
● DN 15 to 300 (½" to 12")
Full bore sensor:
● DN 350 to 1200 (14" to 48")
Full bore sensor:
● DN 25 to 2000 (1" to 78")
Measuring principle
Electromagnetic induction
Excitation frequency
(Mains supply: 50 Hz/60 Hz)
DN 15 to 65 (½" to 2½"):
● 12.5 Hz / 15 Hz
DN 80 to 150 (3" to 6"):
● 6.25 Hz / 7.5 Hz
DN 200 to 300 (8" to 12"):
● 3.125 Hz / 3.75 Hz
DN 350 to 1200 (14" to 48"):
● 1.5625 Hz / 1.875 Hz
DN 25 to 65 (1" to 2½"):
● 12.5 Hz / 15 Hz
DN 80 to 150 (3" to 6"):
● 6.25 Hz / 7.5 Hz
DN 200 to 1200 (8" to 48"):
● 3.125 Hz / 3.75 Hz
DN 1400 to 2000 (54" to 78"):
● 1.5625 Hz / 1.875 Hz
SITRANS FM MAG 5100 W
Operating Instructions, 01/2020, A5E03063678-AC
49