MOBILETT Plus/ Plus E
SPR8-215.898.02
Page 10 of 28
Siemens-Elema AB
Rev. 01
01.02
SPS-UD
Solna, Sweden
10 - 10
Mechanical systems
•
Tighten screws (2) and (3).
•
Tighten the adjustment screw until the arm and single tank are held in the uppermost
position by the spring force alone.
MOBILETT Plus only
The installation of the turn plate is performed in the reverse order.
•
Install the turn plate.
•
Reattach the shafts.
•
Reattach the bearing rings.
•
Reattach the locking rings.
•
Connect the ground cable and K68.
All MOBILETT:s
•
Reinstall the supporting wheels.
•
Remove the ceiling support carefully and check that the single tank remains in the
uppermost position by itself. If not, tighten the adjustment screw.
•
Move the arm into the park position.
•
Lower the unit to the floor and block the rear wheels.
•
Check the arm and single tank movement. If necessary, adjust according to section
"Adjustment of the arm system".
•
Reattach all covers except the system upper cover.
•
Set switch S10 in the ON position to charge the capacitor bank.
•
Reattach the system upper cover.
•
Check the function of the turn plate and the unit by releasing one exposure.