MOBILETT Plus/ Plus E
SPR8-215.898.02
Page 2 of 6
Siemens-Elema AB
Rev. 01
01.02
SPS-UD
Solna, Sweden
8 - 2
Miscellaneous
•
Switch on the unit and check that no messages are displayed. If an Err or CAL message
is displayed, proceed as follows:
•
Change to service mode by switching ON S 1:1 on D1.
•
Perform the adaptation according to section "P10 - Fast adaptation".
•
Set the selected mAs limit, refer to the section "P15 - Upper mAs limit".
•
Set the selected kV step and limit, refer to the section "P14 - Number of kV steps and
upper kV limit".
•
Read out the exposure counter, refer to the section "P16 - Show the exposure counter".
•
Switch off S 1:1 on D1 (OFF).
•
Reattach the system upper cover.
•
Make an exposure at 90 kV, 10 mAs to check that everything is functioning correctly.