Pull out the cable and rewind it to bring the tension back to the cable winder spring.
Repeat that six times.
•
Reinstall upper plate with distance ring, plastic role and Y-plate.
•
Pull out the cable completely.
•
Test cable winder and brake function.
•
Reinstallation is performed in the reverse order.
Fig. 8
The cable winder should not rewind by itself.
The brake force is adjusted via right screw.
nut clockwise -> stronger (pos. 1 / Fig. 8).
nut counterclockwise -> softer.
MOB00620