O P E R A T I N G I N S T R U C T I O N S | ZIRKOR200 Ex-G
Subject to change without notice
COMMISSIONING
6
▸
Is the allocation of the pneumatic connections correct and are the connections gas-
tight? (
see “Pneumatic connections and settings”, page 54
)
▸
Make sure there are not leaks on the analyzer unit - is the counter flange welded gas-
tight to the flue gas duct, are the flange bolts sufficiently tightened? Were flange gaskets
used? (
see “Fitting the analyzer unit”, page 36
)
▸
Do the conditions on site correspond to the specifications in the Data Sheets? (
▸
Does the heating monitoring still have a valid calibration/function check (
shutdown in case of malfunction”, page 81
6.3
Operation
In the Ex-area, the ZIRKOR200 Ex-G can only be operated via SICK REMOTE using the
ZIRKOR Remote App.
The enclosure must be opened to operate the device via push-buttons. The enclosure may
only be opened in an Ex-free environment. The device must not be opened during operation
when it is not possible to ensure freedom from explosion hazards.