O P E R A T I N G I N S T R U C T I O N S | ZIRKOR200 Ex-G
Subject to change without notice
LAYOUT AND FUNCTION
4
Fig. 5: ZIRKOR200 Ex-G Analyzer unit with cooling protection pipe for flue gas temperature up to
1400 °C
1 Control unit / IP66
A
Potentially explosive atmosphere Zone 1 2G IIC T6
Ambient temperatures: -20 °C to +55 °C (-4 °F to +131 °F)
2 In-situ analyzer unit/ IP66
B
Potentially explosive atmosphere Zone 1 2G IIC T3
Flue gas duct / combustion chamber - flue gas temperature up to
1400 °C
3 Pneumatic line
C
Flue gas direction
4 O
2
connection cable, analyzer electronics - control unit
5 Solenoid valve (optional)
6 Counter flange (optional)
F
Space requirements: 2.0 m for straight construction, 0.8 m for 90°
angle construction
7 Insulation: Customer*
8 Duct wall
a
Output signals (analog and digital)
9 Solenoid valve control cable (only with solenoid valve
option)
b
Power supply, power voltage 230 V/115 V ±10% 50 to 60 Hz, max.
400 VA
10 Pneumatic unit IP 66
c
Test gas inlet
11 Connection cable, pneumatic unit – control unit (only
with solenoid valve option)
d
Instrument air inlet
12 Connection cable
13 Connection cable
* Note: Risk of corrosion due to lack of insulation