Operating Instructions
Chapter
7
RFH620 Interrogator
Commissioning and configuration
8013105/0000/2009-05-12
©
SICK AG · Division Auto Ident · Germany · All rights reserved
55
7.5
Default setting
The values of the default setting are permanently saved in the interrogator (ROM) and in the
database of the SOPAS-ET configuration software in the device-specific jar file (see
chapter 7.4 Initial commissioning, page 52
). A PC is not required to start up the interrogator
with the default setting.
7.5.1
Resetting the default setting in the interrogator
Prerequisite
The SOPAS-ET configuration software is connected online to the interrogator.
Two default setting types can be called up via the SOPAS-ET configuration software:
•
Complete default setting (
LOAD
DEFAULT
SETTING
)
SOPAS-ET resets all parameter values of the interrogator to default. Settings, which
have been previously made for the communication parameters of the Ethernet inter-
faces or serial data interfaces (e.g. Ethernet address), are overwritten. In the process,
the connection(s) to the interrogator disappear(s) as the case may be and must to be
reconfigured.
•
Application-specific default setting (
LOAD
APPLICATION
DEFAULT
SETTING
)
SOPAS-ET resets the parameter values of the interrogator, but does not change the
communication parameters. Settings, which have been previously made for the com-
munication parameters of the Ethernet interfaces or serial data interfaces, are kept
and the current connection(s) to the interrogator remain(s) established.
1. In order to discard changes to the parameter set as described above, select the cor-
responding command in the menu bar under RFH620.
The SOPAS-ET configuration software loads the default setting from the interrogator
and displays the parameter values in the register tabs. In the interrogator, the default
setting will first be active in the temporary main memory only.
The default setting can also be saved in the PC or printed (see
play and print the current parameter set, page 54
).
2. In the menu bar under
TOOLS
, select the command
LOGIN
TO
DEVICE
.
3. In the dialog window under
USERLEVEL
in the list box, select the entry
AUTHORISED
CLIENT
.
If the parameter set is password-protected, enter the password "client" in P
ASSWORD
.
4. Click on OK to confirm the dialog window.
5. In the menu bar under RFH620, select the command
PARAMETER
/
SAVE
PERMANENT
.
The SOPAS-ET configuration software transfers the default setting to the permanent pa-
rameter memory (PROM) of the interrogator.
If the interrogator is equipped with the Micro SD memory card or connected to a con-
nection module CDB620/CDM420 with parameter memory module CMC600, the de-
fault setting will be permanently transferred to the parameter memory CMC600 and
memory card respectively.
Important
Once the default setting has been restored, password-protection is deactivated.